8 ml Clear Screw Vials, 15-425 Thread, 16.6 × 60 mm, 100/pk
8 ml, 15 mm, Clear, 16.6 × 60 mm
Lead time 3-4 weeks, ships from the Czech Republic (EU)
8 ml (16.6 × 60 mm) and 12 ml (18.5 × 65 mm) glass screw vials with a 15-425 thread, in clear and amber glass.
They suit larger samples, standards and storage, in packs of 100. Items not in stock have a lead time of 3 to 4 weeks.
8 ml, 15 mm, Clear, 16.6 × 60 mm
Lead time 3-4 weeks, ships from the Czech Republic (EU)
12 ml, 15 mm, Clear, 18.5 × 65 mm
Lead time 3-4 weeks, ships from the Czech Republic (EU)
8 ml, 15 mm, Amber, 16.6 × 60 mm
Lead time 3-4 weeks, ships from the Czech Republic (EU)
12 ml, 15 mm, Amber, 18.5 × 65 mm
Lead time 3-4 weeks, ships from the Czech Republic (EU)
No products match these filters.
Start from your autosampler: it decides the vial size and neck. Then pick the glass colour and the cap.
| Neck | Closure | Typical use |
|---|---|---|
| 9 mm short thread (9-425) | Screw cap with septum | The most common HPLC format; wide opening, easy to fill |
| 8 mm (8-425) | Screw cap with septum | Older and some specific autosamplers |
| 10 mm (10-425) | Screw cap with septum | Specific autosampler models |
| 11 mm crimp or snap | Aluminium crimp cap or snap cap | GC and volatile samples (crimp seals tightest) |
| 18 mm and 20 mm | Magnetic screw or crimp cap | Headspace GC |
Standard 2 ml vials measure 11.6 × 32 mm (often written 12 × 32 mm). Check the vial format in your autosampler manual; a free sample lets you test the fit in your rack.
Use amber glass for light-sensitive compounds. Use clear glass for everything else, so you can see the fill level and any particles.
PTFE/silicone septa suit most HPLC and GC work and reseal for repeat injections. Pre-slit septa reduce needle wear and stop a vacuum forming in the vial. Order caps and vials together: on EROHE, equal packs of 9 mm vials and caps get 5% off both.
For volumes under 1 ml, use a glass insert in a standard vial, or a vial with a built-in conical insert, so the needle can still reach the sample.
15 mm screw caps (ND15), available with a centre hole or closed top and PTFE/silicone septa.
Choose by the volume you need to hold plus headspace, and by the height your rack or instrument takes: the 8 ml vial is 60 mm tall and the 12 ml vial 65 mm.
Only on autosamplers or samplers that accept vials of these dimensions; check your instrument's vial specification.